Pengaruh Getaran Centrifuge Terhadap Hasil Pemeriksaan Laju Endap Darah (LED)
(1) Akademi Analis Kesehatan Borneo Lestari
(*) Corresponding Author
Abstract
The Erythrocyte Sedimentation Rate (LED) is a hematological examination which is a preliminary examination in establishing a patient's diagnosis. Many LED tests have factors that can affect the results, one of them is vibration. In some health service units, LED checks are carried out close to the Centrifuge, which has vibrations, for reasons of room efficiency. This study aims to determine the effect of Centrifuge vibration on the results of the Westergren method LED examination. The study was conducted with 6 repetitions. This type of research used is quasi-experimental with static group comparison design. The research obtained is presented in table form, followed by ANOVA test. ANOVA test gives a significant value of 0.305 which indicates that there is no significant effect by the vibration of the Centrifuge on the LED results with variations in the distance of 0.5m, 1m, and 1.5m
Full Text:
PDFArticle Metrics
Abstract view : 896 timesPDF - 331 times
DOI: https://doi.org/10.26714/jlabmed.3.1.2019.10-14
Refbacks
- There are currently no refbacks.
Copyright (c) 2019 Jurnal Labora Medika
View My Stats
Program Studi Teknologi Laboratorium Medik Universitas Muhammadiyah Semarang
Jl. Kedungmundu Raya No. 18 Semarang Gedung Laboratorium Kesehatan Universitas Muhammadiyah Semarang
Phone: 02476740295
Email: jlabmed@unimus.ac.id